Right ventricular lead perforation is a rare but serious and potentially life-threatening complication of pacemaker or defibrillator lead implantation. This report describes a patient with Brugada syndrome in whom the diagnosis of asymptomatic right ventricular perforation by an implantable cardioverter defibrillator lead was detected 12 days after implantation, thanks to a report from home monitoring system. The patient was admitted to our institution, where the lead was explanted and replaced. This case illustrates the potential lifesaving benefit of the home monitoring system in patients with implantable cardioverter defibrillator.

Asymptomatic right ventricular perforation by an implantable cardioverterdefibrillator lead detected by home monitoring system

MIGLIORE, FEDERICO;TARANTINI, GIUSEPPE;BUJA, GIANFRANCO;ILICETO, SABINO;CORRADO, DOMENICO
2010

Abstract

Right ventricular lead perforation is a rare but serious and potentially life-threatening complication of pacemaker or defibrillator lead implantation. This report describes a patient with Brugada syndrome in whom the diagnosis of asymptomatic right ventricular perforation by an implantable cardioverter defibrillator lead was detected 12 days after implantation, thanks to a report from home monitoring system. The patient was admitted to our institution, where the lead was explanted and replaced. This case illustrates the potential lifesaving benefit of the home monitoring system in patients with implantable cardioverter defibrillator.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/2432497
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 8
social impact