Abstract Background: The detection of faecal occult blood is a fundamental step in making an early diagnosis of colorectal cancer. The aim of the present study was to evaluate the stability of haemoglobin in faeces collected with two sampling devices specific for faecal immunochemical testing (FOB Gold Tube Screen and FOB Gold Tube NG) that contain different preservative buffers (buffer H, BH, and buffer N, BN, respectively). Methods: Fifteen true positive faecal samples were collected with both devices. A pool from each sample was made. Each pool was portioned and stored at +4°C, +21°C and +32°C for 10 days. One aliquot of each pool stored at each of the respective temperatures was tested at five time intervals between sampling and analysis. The same procedure was followed for three synthetic haemoglobin solutions in both buffers. Results: The percentage of cumulative faecal haemoglobin decrease (HbCD%) was evaluated. No significant difference was found between BH and BN in HbCD% at +4°C (p=0.106); at +21°C and +32°C, HbCD% was lower in BH than in BN samples (p=0.002 and p

A new sampling device for faecal immunochemical testing: haemoglobin stability is still an open issue

GNATTA, ELISA;ZANINOTTO, MARTINA;PADOAN, ANDREA;GJINI, ROMELDA;PLEBANI, MARIO
2014

Abstract

Abstract Background: The detection of faecal occult blood is a fundamental step in making an early diagnosis of colorectal cancer. The aim of the present study was to evaluate the stability of haemoglobin in faeces collected with two sampling devices specific for faecal immunochemical testing (FOB Gold Tube Screen and FOB Gold Tube NG) that contain different preservative buffers (buffer H, BH, and buffer N, BN, respectively). Methods: Fifteen true positive faecal samples were collected with both devices. A pool from each sample was made. Each pool was portioned and stored at +4°C, +21°C and +32°C for 10 days. One aliquot of each pool stored at each of the respective temperatures was tested at five time intervals between sampling and analysis. The same procedure was followed for three synthetic haemoglobin solutions in both buffers. Results: The percentage of cumulative faecal haemoglobin decrease (HbCD%) was evaluated. No significant difference was found between BH and BN in HbCD% at +4°C (p=0.106); at +21°C and +32°C, HbCD% was lower in BH than in BN samples (p=0.002 and p
File in questo prodotto:
File Dimensione Formato  
Gnatta et al, 2014.pdf

accesso aperto

Tipologia: Published (publisher's version)
Licenza: Accesso libero
Dimensione 545.38 kB
Formato Adobe PDF
545.38 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/2827257
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 7
social impact