This work investigates capabilities and accuracy of metrological X-ray Computed Tomography (CT) in quantifying wear volumes and characterizing the geometry of prosthetic joint components. Advantages and limitations are determined in comparison with measurements by Coordinate Measuring Machines (CMMs). The work demonstrates that CT is advantageous especially for measuring polymeric prosthetic components.

X-ray computed tomography for wear measurement of prosthetic components

CARMIGNATO, SIMONE;ZANINI, FILIPPO;BALCON, MANUEL;SAVIO, ENRICO
2014

Abstract

This work investigates capabilities and accuracy of metrological X-ray Computed Tomography (CT) in quantifying wear volumes and characterizing the geometry of prosthetic joint components. Advantages and limitations are determined in comparison with measurements by Coordinate Measuring Machines (CMMs). The work demonstrates that CT is advantageous especially for measuring polymeric prosthetic components.
2014
Conference Proceedings - 14th International Conference of the European Society for Precision Engineering and Nanotechnology, EUSPEN 2014
9780956679031
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3155129
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? ND
social impact