Background: At the beginning of 2020, a high number of COVID-19 cases affected Italy in a short period, causing a difficult supply of medical equipment. To deal with the problem, many healthcare operators readapted different masks to medical devices, but no experiment was conducted to evaluate their performance. The aims of our study were: to assess the performances of three masks and a CPAP helmet in their original configuration and after modifications, in the maintenance of mean pressure and half-amplitude variations using different PEEP valves and to analyse the impact of antibacterial (AB) or antibacterial-viral (ABV) pre-valve PEEP filters on the effective PEEP delivered to the patients. Four pressure ports were installed on each mask (three on helmet), mean values and half amplitudes of pressure were recorded. Tests were performed with any, AB, ABV filter before the PEEP valve. CPAP helmet was the most efficient interface producing more stable mean pressure and less prominent half-amplitude variations but the non-medical masks, especially after the modifications, maintained a stable mean pressure value with only a moderate increase of half-amplitude. The use of AB and ABV filters, produced respectively an increase of 2,23% and 16.5% in mean pressure, compared to no filter condition. CPAP helmet is the most reliable interface in terms of detected performance, but readapted masks can assure almost a similar performance. The use of ABV filters before the PEEP valve significantly increases the detected mean pressure while the AB filters have almost a neutral effect.

Performance assessment of medical and non-medical CPAP interfaces used during the COVID-19 pandemic

Lorenzo Capponi;Gianluca Rossi;
2022

Abstract

Background: At the beginning of 2020, a high number of COVID-19 cases affected Italy in a short period, causing a difficult supply of medical equipment. To deal with the problem, many healthcare operators readapted different masks to medical devices, but no experiment was conducted to evaluate their performance. The aims of our study were: to assess the performances of three masks and a CPAP helmet in their original configuration and after modifications, in the maintenance of mean pressure and half-amplitude variations using different PEEP valves and to analyse the impact of antibacterial (AB) or antibacterial-viral (ABV) pre-valve PEEP filters on the effective PEEP delivered to the patients. Four pressure ports were installed on each mask (three on helmet), mean values and half amplitudes of pressure were recorded. Tests were performed with any, AB, ABV filter before the PEEP valve. CPAP helmet was the most efficient interface producing more stable mean pressure and less prominent half-amplitude variations but the non-medical masks, especially after the modifications, maintained a stable mean pressure value with only a moderate increase of half-amplitude. The use of AB and ABV filters, produced respectively an increase of 2,23% and 16.5% in mean pressure, compared to no filter condition. CPAP helmet is the most reliable interface in terms of detected performance, but readapted masks can assure almost a similar performance. The use of ABV filters before the PEEP valve significantly increases the detected mean pressure while the AB filters have almost a neutral effect.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3448478
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact