A new method for the calibration of a superconducting gravity meter is described, in which a 273 Kg annular mass is placed around the meter and is moved up and down. The geometry of the apparatus is easy to model and the accuracy in the computation of the gravity variation induced by the mass, 6.7 μgal, is limited only by the accuracy in the knowledge of value of the gravitational constant. Measurements done in '91 and '92 for the calibration of the instrument GWR-T015 are described. The calibration factor has been determined with a precision of about 0.3%. © 1995 Springer-Verlag.

A CALIBRATION SYSTEM FOR SUPERCONDUCTING GRAVIMETERS

ACHILLI, VLADIMIRO;
1995

Abstract

A new method for the calibration of a superconducting gravity meter is described, in which a 273 Kg annular mass is placed around the meter and is moved up and down. The geometry of the apparatus is easy to model and the accuracy in the computation of the gravity variation induced by the mass, 6.7 μgal, is limited only by the accuracy in the knowledge of value of the gravitational constant. Measurements done in '91 and '92 for the calibration of the instrument GWR-T015 are described. The calibration factor has been determined with a precision of about 0.3%. © 1995 Springer-Verlag.
1995