Background: We organized and conducted a proficiency testing study (CardioOrmocheck) to evaluate the differences in analytical performance of brain natriuretic peptide (BNP) and N-terminal proBNP (NT-proBNP) immunoassays. Methods: Approximately 90 Italian laboratories were involved in the 2005-2007 proficiency testing cycles, while 112 laboratories took part in the 2008 cycle (from January to May 2008). A total of 28 study samples were measured by participating laboratories for a total of 2354 determinations. Results: The mean total variability for BNP (50.6%CV) was significantly higher than that for NT-proBNP (8.4%CV). In addition, the mean variability due to differences between-methods (46.4%CV) comprised the majority of the total variability for BNP. Between-method variability for BNP comprised, on average, 84% of total variability, while the within-method variability comprised an average of 20.2%CV. On the contrary, for NT-proBNP the within-method variability (7.3%CV) represented the majority of total variability (average 75%), while between-method variability was smaller (4.1%CV). Imprecision around the cut-off value showed marked differences among methods, especially for BNP immunoassay methods. In addition, BNP methods were affected by large systematic differences, for example an average 2.7-fold difference between Access and ADVIA Centaur methods, while agreement between NT-proBNP methods was better (an average 1.2-fold difference between Dimension and ECLIA on the Elecsys methods). Conclusions: This multicenter collaborative study demonstrates that there are significant differences in analytical characteristics and measured values among the most popular commercial methods for BNP and NT-proBNP. Clinicians should be very careful when comparing results obtained by laboratories that use different methods. Clin Chem Lab Med 2009;47:762-8.

Proficiency testing project for brain natriuretic peptide (BNP) and the N-terminal part of the propeptide of BNP (NT-proBNP) immunoassays: the CardioOrmocheck study.

PLEBANI, MARIO;
2009

Abstract

Background: We organized and conducted a proficiency testing study (CardioOrmocheck) to evaluate the differences in analytical performance of brain natriuretic peptide (BNP) and N-terminal proBNP (NT-proBNP) immunoassays. Methods: Approximately 90 Italian laboratories were involved in the 2005-2007 proficiency testing cycles, while 112 laboratories took part in the 2008 cycle (from January to May 2008). A total of 28 study samples were measured by participating laboratories for a total of 2354 determinations. Results: The mean total variability for BNP (50.6%CV) was significantly higher than that for NT-proBNP (8.4%CV). In addition, the mean variability due to differences between-methods (46.4%CV) comprised the majority of the total variability for BNP. Between-method variability for BNP comprised, on average, 84% of total variability, while the within-method variability comprised an average of 20.2%CV. On the contrary, for NT-proBNP the within-method variability (7.3%CV) represented the majority of total variability (average 75%), while between-method variability was smaller (4.1%CV). Imprecision around the cut-off value showed marked differences among methods, especially for BNP immunoassay methods. In addition, BNP methods were affected by large systematic differences, for example an average 2.7-fold difference between Access and ADVIA Centaur methods, while agreement between NT-proBNP methods was better (an average 1.2-fold difference between Dimension and ECLIA on the Elecsys methods). Conclusions: This multicenter collaborative study demonstrates that there are significant differences in analytical characteristics and measured values among the most popular commercial methods for BNP and NT-proBNP. Clinicians should be very careful when comparing results obtained by laboratories that use different methods. Clin Chem Lab Med 2009;47:762-8.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/2379845
Citazioni
  • ???jsp.display-item.citation.pmc??? 6
  • Scopus 63
  • ???jsp.display-item.citation.isi??? 63
social impact