We report on the performances of CsI(Tl) crystals as gamma -ray detectors, compared with standard NaI(Tl). Energy resolution from E-gamma = 122 keV up to E-gamma = 7.6 MeV and light output yield of different CsI(Tl) crystals and read-out configurations (photomultiplier tubes, photodiodes and avalanche photodiodes) have been measured.
CsI(Tl) scintillators as g-ray detectors for the identification of hidden explosives
VIESTI, GIUSEPPE;LUNARDON, MARCELLO;
2001
Abstract
We report on the performances of CsI(Tl) crystals as gamma -ray detectors, compared with standard NaI(Tl). Energy resolution from E-gamma = 122 keV up to E-gamma = 7.6 MeV and light output yield of different CsI(Tl) crystals and read-out configurations (photomultiplier tubes, photodiodes and avalanche photodiodes) have been measured.File in questo prodotto: