OBJECTIVES: : Although the upper arm has the shape of a truncated cone, cylindrical cuffs and bladders are currently used for blood pressure (BP) measurement. The aim of this study was to ascertain whether cylindrical and tronco-conical cuffs provide different readings according to arm size and shape. DESIGN: : We studied 220 individuals with arm circumference ranging from 22 to 42.5 cm. Four different cylindrical and four different tronco-conical bladders of appropriate size were used. Sequential same-arm measurements were performed in triplicate by two observers using the two cuffs in a random order. In 100 individuals, the actual pressure transmitted to the arm surface by the two cuffs at the central point was also measured. RESULTS: : Upper arm shape was tronco-conical in all of the individuals. In a multiple regression, conicity was related to arm circumference (P < 0.001) and length (P = 0.001). Arm conicity and size were independently related to the between-cuff SBP (P = 0.001 and 0.002, respectively) and DBP (P = 0.001 and <0.001, respectively) discrepancies. In the group with arm circumference of 37.5-42.5 cm, the cylindrical cuff overestimated BP measured with the tronco-conical cuff by 2.0 ± 0.4/1.8 ± 0.3 mmHg (P = 0.001 and <0.001). In this group, 15% of individuals found hypertensive with the cylindrical cuff were normotensive when assessed with the conical cuff. Differences as great as 9.7/7.8 mmHg were found in individuals with large arms and slant angle equal to or less than 83°. CONCLUSION: : In obese people, the upper arm may have a pronounced tronco-conical shape and cylindrical cuffs may overestimate BP. Tronco-conical cuffs should be used for BP measurement in individuals with large arms.

Rectangular cuffs may overestimate blood pressure in individuals with large conical arms.

PALATINI, PAOLO;FANIA, CLAUDIO;
2012

Abstract

OBJECTIVES: : Although the upper arm has the shape of a truncated cone, cylindrical cuffs and bladders are currently used for blood pressure (BP) measurement. The aim of this study was to ascertain whether cylindrical and tronco-conical cuffs provide different readings according to arm size and shape. DESIGN: : We studied 220 individuals with arm circumference ranging from 22 to 42.5 cm. Four different cylindrical and four different tronco-conical bladders of appropriate size were used. Sequential same-arm measurements were performed in triplicate by two observers using the two cuffs in a random order. In 100 individuals, the actual pressure transmitted to the arm surface by the two cuffs at the central point was also measured. RESULTS: : Upper arm shape was tronco-conical in all of the individuals. In a multiple regression, conicity was related to arm circumference (P < 0.001) and length (P = 0.001). Arm conicity and size were independently related to the between-cuff SBP (P = 0.001 and 0.002, respectively) and DBP (P = 0.001 and <0.001, respectively) discrepancies. In the group with arm circumference of 37.5-42.5 cm, the cylindrical cuff overestimated BP measured with the tronco-conical cuff by 2.0 ± 0.4/1.8 ± 0.3 mmHg (P = 0.001 and <0.001). In this group, 15% of individuals found hypertensive with the cylindrical cuff were normotensive when assessed with the conical cuff. Differences as great as 9.7/7.8 mmHg were found in individuals with large arms and slant angle equal to or less than 83°. CONCLUSION: : In obese people, the upper arm may have a pronounced tronco-conical shape and cylindrical cuffs may overestimate BP. Tronco-conical cuffs should be used for BP measurement in individuals with large arms.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/2489686
Citazioni
  • ???jsp.display-item.citation.pmc??? 9
  • Scopus 44
  • ???jsp.display-item.citation.isi??? ND
social impact