The Sorin Pericarbon Freedom (SPF) valve is a stentless bioprosthesis made from bovine pericardium, with a peculiar design aimed at preventing the mechanical failures observed with old models of stented pericardial bioprostheses. Herein, the case is described of a patient who presented with severe regurgitation of a SPF six years after aortic valve replacement, caused by commissural dehiscence. Both, microradiographic and histologic investigations, revealed mild calcific deposits and massive lipid infiltration, thus confirming that a patient-related mechanism such as 'atheromasia' can account for structural valve deterioration also in recipients of pericardial bioprostheses.

Lipid insudation as a cause of structural failure of a stentless pericardial bioprosthesis.

DELLA BARBERA, MILA;VALENTE, MARIALUISA;THIENE, GAETANO;
2012

Abstract

The Sorin Pericarbon Freedom (SPF) valve is a stentless bioprosthesis made from bovine pericardium, with a peculiar design aimed at preventing the mechanical failures observed with old models of stented pericardial bioprostheses. Herein, the case is described of a patient who presented with severe regurgitation of a SPF six years after aortic valve replacement, caused by commissural dehiscence. Both, microradiographic and histologic investigations, revealed mild calcific deposits and massive lipid infiltration, thus confirming that a patient-related mechanism such as 'atheromasia' can account for structural valve deterioration also in recipients of pericardial bioprostheses.
2012
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/2837083
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 2
social impact