BACKGROUND: The SOURCE 3 Registry (SAPIEN Aortic Bioprosthesis European Outcome) is a European multicenter, observational registry of the latest generation of transcatheter heart valve, the SAPIEN 3 (Edwards Lifesciences, Irvine, CA). Its purpose is to document outcomes of clinical safety and performance after European approval was given. METHODS: Here, we present the 30-day outcome of the SOURCE 3 Registry. All data are self-reported, and all participating centers have committed to support their consecutive experience with the SAPIEN 3 transcatheter heart valve, dependent on patient consent, before the start of the study. Adverse events are defined with Valve Academic Research Consortium 2 criteria and adjudicated by an independent clinical events committee. RESULTS: A total of 1950 patients from 80 centers in 10 countries were enrolled between July 2014 and October 2015. Of those, 1947 patients underwent transcatheter aortic valve implantation (TAVI) with the SAPIEN 3 (mean age, 81.6±6.6 years; 48.1% female). Main comorbidities included coronary artery disease (51.5%), renal insufficiency (27.4%), diabetes mellitus (29.5%), chronic obstructive pulmonary disease (16.0%), and a mean logistic EuroSCORE of 18.3±13.2. Transfemoral access was used in 87.1% (n=1695); nontransfemoral, in 252 patients. Conscious sedation was used in 59.9% of transfemoral procedures, and in 50% of patients, TAVI was performed without aortic balloon valvuloplasty. Implantation success (1 valve in the intended location) was 98.3%. Conversion to conventional surgery (0.6%) and use of cardiopulmonary bypass (0.7%) were rare. Adverse events were low, with site-reported 30-day all-cause mortality of 2.2%, cardiovascular mortality of 1.1%, stroke of 1.4%, major vascular complications of 4.1%, life-threatening bleeding of 5%, and post-TAVI pacemaker implantation of 12%. Moderate or greater paravalvular regurgitation was observed in 3.1% of reporting patients. CONCLUSIONS: Results from the SOURCE 3 Registry demonstrate contemporary European trends and good outcomes of TAVI in daily practice when this third-generation TAVI device is used.

SOURCE 3 Registry: Design and 30-day results of the european postapproval registry of the latest generation of the sapien 3 transcatheter heart valve

Tarantini, Giuseppe;
2017

Abstract

BACKGROUND: The SOURCE 3 Registry (SAPIEN Aortic Bioprosthesis European Outcome) is a European multicenter, observational registry of the latest generation of transcatheter heart valve, the SAPIEN 3 (Edwards Lifesciences, Irvine, CA). Its purpose is to document outcomes of clinical safety and performance after European approval was given. METHODS: Here, we present the 30-day outcome of the SOURCE 3 Registry. All data are self-reported, and all participating centers have committed to support their consecutive experience with the SAPIEN 3 transcatheter heart valve, dependent on patient consent, before the start of the study. Adverse events are defined with Valve Academic Research Consortium 2 criteria and adjudicated by an independent clinical events committee. RESULTS: A total of 1950 patients from 80 centers in 10 countries were enrolled between July 2014 and October 2015. Of those, 1947 patients underwent transcatheter aortic valve implantation (TAVI) with the SAPIEN 3 (mean age, 81.6±6.6 years; 48.1% female). Main comorbidities included coronary artery disease (51.5%), renal insufficiency (27.4%), diabetes mellitus (29.5%), chronic obstructive pulmonary disease (16.0%), and a mean logistic EuroSCORE of 18.3±13.2. Transfemoral access was used in 87.1% (n=1695); nontransfemoral, in 252 patients. Conscious sedation was used in 59.9% of transfemoral procedures, and in 50% of patients, TAVI was performed without aortic balloon valvuloplasty. Implantation success (1 valve in the intended location) was 98.3%. Conversion to conventional surgery (0.6%) and use of cardiopulmonary bypass (0.7%) were rare. Adverse events were low, with site-reported 30-day all-cause mortality of 2.2%, cardiovascular mortality of 1.1%, stroke of 1.4%, major vascular complications of 4.1%, life-threatening bleeding of 5%, and post-TAVI pacemaker implantation of 12%. Moderate or greater paravalvular regurgitation was observed in 3.1% of reporting patients. CONCLUSIONS: Results from the SOURCE 3 Registry demonstrate contemporary European trends and good outcomes of TAVI in daily practice when this third-generation TAVI device is used.
2017
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3243429
Citazioni
  • ???jsp.display-item.citation.pmc??? 35
  • Scopus 168
  • ???jsp.display-item.citation.isi??? 160
social impact