Background NeoTapAdvancedSupport (NeoTapAS) is a mobile application, based on a screen tapping method that calculates the heart rate (HR). We aimed to evaluate the accuracy of NeoTapAS in reliably determining HR from auscultation in a high-fidelity simulated newborn resuscitation scenario. Methods Paediatric residents assessed HR by auscultation plus NeoTapAS in an asphyxiated term infant scenario and orally communicated the estimated HR. An external observer simultaneously documented the actual HR set in the manikin and the communicated HR. Results One hundred and sixty HR measurements were recorded. The agreement between communicated and set HR was good (Cohen's kappa 0.80, 95% CI 0.72 to 0.87; Bangdiwala's weighted agreement strength statistic 0.93). Bland-Altman plot showed a mean difference between communicated and set HR values of 1 beats per minute (bpm) (95% agreement limits -9 to 11 bpm). Conclusion NeoTapAS showed a good accuracy in estimating HR and it could be an important resource in settings with limited availability of ECG monitor.
Heart rate assessment using NeoTapAdvancedSupport: A simulation study
Cavallin, Francesco;Trevisanuto, Daniele
2019
Abstract
Background NeoTapAdvancedSupport (NeoTapAS) is a mobile application, based on a screen tapping method that calculates the heart rate (HR). We aimed to evaluate the accuracy of NeoTapAS in reliably determining HR from auscultation in a high-fidelity simulated newborn resuscitation scenario. Methods Paediatric residents assessed HR by auscultation plus NeoTapAS in an asphyxiated term infant scenario and orally communicated the estimated HR. An external observer simultaneously documented the actual HR set in the manikin and the communicated HR. Results One hundred and sixty HR measurements were recorded. The agreement between communicated and set HR was good (Cohen's kappa 0.80, 95% CI 0.72 to 0.87; Bangdiwala's weighted agreement strength statistic 0.93). Bland-Altman plot showed a mean difference between communicated and set HR values of 1 beats per minute (bpm) (95% agreement limits -9 to 11 bpm). Conclusion NeoTapAS showed a good accuracy in estimating HR and it could be an important resource in settings with limited availability of ECG monitor.Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.