Clotting of the filters in continuous arterio-venous hemofiltration (CAVH) has been frequently related to filtration pressure equilibrium (FPE). The geometry of the filter and the blood flow, together with filtration fraction, are the factors affecting FPE. A new method based on scintigraphic imaging of the filter is described to demonstrate FPE and identify the operational conditions influencing it. A radiolabelled marker molecule (albumin macroaggregates + Tc99) is added to the blood circulating at various flows through the filter. Changes in the concentration of this molecule detected by a gamma camera are used to calculate the water fluxes in different sections of the filter. When the curve reaches a plateau, no further changes occur and FPE is achieved. The study used Amicon D-30 hemofilters and FPE points were achieved for blood flows of 75, 110 and 150 ml/min respectively 8 cm, 12 cm and 18 cm from the inlet. The study not only demonstrates the occurrence of FPE, but also identifies the length of the filter required at a given blood flow to avoid FPE (achieving less clotting and lower heparin requirement). The Amicon diafilter family, having filters of different length, is useful for "personalizing" the filter in CAVH.

ULTRAFILTRATION AND PRESSURE PROFILES IN CONTINUOUS ARTERIOVENOUS HEMOFILTRATION STUDIES BY COMPUTERIZED SCINTIGRAPHIC IMAGING

RONCO C;
1991

Abstract

Clotting of the filters in continuous arterio-venous hemofiltration (CAVH) has been frequently related to filtration pressure equilibrium (FPE). The geometry of the filter and the blood flow, together with filtration fraction, are the factors affecting FPE. A new method based on scintigraphic imaging of the filter is described to demonstrate FPE and identify the operational conditions influencing it. A radiolabelled marker molecule (albumin macroaggregates + Tc99) is added to the blood circulating at various flows through the filter. Changes in the concentration of this molecule detected by a gamma camera are used to calculate the water fluxes in different sections of the filter. When the curve reaches a plateau, no further changes occur and FPE is achieved. The study used Amicon D-30 hemofilters and FPE points were achieved for blood flows of 75, 110 and 150 ml/min respectively 8 cm, 12 cm and 18 cm from the inlet. The study not only demonstrates the occurrence of FPE, but also identifies the length of the filter required at a given blood flow to avoid FPE (achieving less clotting and lower heparin requirement). The Amicon diafilter family, having filters of different length, is useful for "personalizing" the filter in CAVH.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3293581
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? 7
social impact