Background: One of the most common adverse events during continuous renal replacement therapy (CRRT) is hypothermia, reported to occur in over 4/10 cases. In turn, hypothermia is known to be associated with higher mortality rates among patients treated in intensive care units (ICU). The present study examined if a novel warming device in the current generation of CRRT systems could lower incidence of hypothermia compared to previous generation technology. Methods: We included ICU patients >18 years, at Skåne University Hospital, Lund from November 2006 to August 2019 and treated with CRRT. Temperature measurements were recorded from the CRRT systems and from the patients hourly. Results: In total, 310 patients treated with the older system vs 32 patients treated using the newer CRRT system were included. We found that historic Prismaflex patients spent 11.43% of their time in hypothermia, as compared to the novel Prismax CRRT system, where 10.06% of patient hours were below 36.0°C (Chi...

Hypothermia During CRRT, a Comparative Analysis

Claudio Ronco;
2020

Abstract

Background: One of the most common adverse events during continuous renal replacement therapy (CRRT) is hypothermia, reported to occur in over 4/10 cases. In turn, hypothermia is known to be associated with higher mortality rates among patients treated in intensive care units (ICU). The present study examined if a novel warming device in the current generation of CRRT systems could lower incidence of hypothermia compared to previous generation technology. Methods: We included ICU patients >18 years, at Skåne University Hospital, Lund from November 2006 to August 2019 and treated with CRRT. Temperature measurements were recorded from the CRRT systems and from the patients hourly. Results: In total, 310 patients treated with the older system vs 32 patients treated using the newer CRRT system were included. We found that historic Prismaflex patients spent 11.43% of their time in hypothermia, as compared to the novel Prismax CRRT system, where 10.06% of patient hours were below 36.0°C (Chi...
2020
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3342538
Citazioni
  • ???jsp.display-item.citation.pmc??? 9
  • Scopus 21
  • ???jsp.display-item.citation.isi??? 18
  • OpenAlex 23
social impact