AIMS: The traditional technique for subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation involves three incisions and a subcutaneous pocket. Recently, a two-incision and intermuscular (IM) technique has been adopted. The PRAETORIAN score is a chest radiograph-based tool that predicts S-ICD conversion testing. We assessed whether the S-ICD implantation technique affects optimal position of the defibrillation system according to the PRAETORIAN score. METHODS AND RESULTS: We analysed consecutive patients undergoing S-ICD implantation. The χ2 test and regression analysis were used to determine the association between the PRAETORIAN score and implantation technique. Two hundred and thirteen patients were enrolled. The S-ICD generator was positioned in an IM pocket in 174 patients (81.7%) and the two-incision approach was adopted in 199 (93.4%). According to the PRAETORIAN score, the risk of conversion failure was classified as low in 198 patients (93.0%), intermediate in 13 (6.1%), and high in 2 (0.9%). Patients undergoing the two-incision and IM technique were more likely to have a low (<90) PRAETORIAN score than those undergoing the three-incision and subcutaneous technique (two-incision: 94.0% vs. three-incision: 78.6%; P = 0.004 and IM: 96.0% vs. subcutaneous: 79.5%; P = 0.001). Intermuscular plus two-incision technique was associated with a low-risk PRAETORIAN score (hazard ratio 3.76; 95% confidence interval 1.01-14.02; P = 0.04). Shock impedance was lower in PRAETORIAN low-risk patients than in intermediate-/high-risk categories (66 vs. 96 Ohm; P = 0.001). The PRAETORIAN score did not predict shock failure at 65 J. CONCLUSION: In this cohort of S-ICD recipients, combining the two-incision technique and IM generator implantation yielded the lowest PRAETORIAN score values, indicating optimal defibrillation system position. CLINICAL TRIAL REGISTRATION: http://clinicaltrials.gov/ Identifier: NCT02275637.

Implantation technique and optimal subcutaneous defibrillator chest position: a PRAETORIAN score-based study

Migliore F.;Giorgi D.;
2020

Abstract

AIMS: The traditional technique for subcutaneous implantable cardioverter-defibrillator (S-ICD) implantation involves three incisions and a subcutaneous pocket. Recently, a two-incision and intermuscular (IM) technique has been adopted. The PRAETORIAN score is a chest radiograph-based tool that predicts S-ICD conversion testing. We assessed whether the S-ICD implantation technique affects optimal position of the defibrillation system according to the PRAETORIAN score. METHODS AND RESULTS: We analysed consecutive patients undergoing S-ICD implantation. The χ2 test and regression analysis were used to determine the association between the PRAETORIAN score and implantation technique. Two hundred and thirteen patients were enrolled. The S-ICD generator was positioned in an IM pocket in 174 patients (81.7%) and the two-incision approach was adopted in 199 (93.4%). According to the PRAETORIAN score, the risk of conversion failure was classified as low in 198 patients (93.0%), intermediate in 13 (6.1%), and high in 2 (0.9%). Patients undergoing the two-incision and IM technique were more likely to have a low (<90) PRAETORIAN score than those undergoing the three-incision and subcutaneous technique (two-incision: 94.0% vs. three-incision: 78.6%; P = 0.004 and IM: 96.0% vs. subcutaneous: 79.5%; P = 0.001). Intermuscular plus two-incision technique was associated with a low-risk PRAETORIAN score (hazard ratio 3.76; 95% confidence interval 1.01-14.02; P = 0.04). Shock impedance was lower in PRAETORIAN low-risk patients than in intermediate-/high-risk categories (66 vs. 96 Ohm; P = 0.001). The PRAETORIAN score did not predict shock failure at 65 J. CONCLUSION: In this cohort of S-ICD recipients, combining the two-incision technique and IM generator implantation yielded the lowest PRAETORIAN score values, indicating optimal defibrillation system position. CLINICAL TRIAL REGISTRATION: http://clinicaltrials.gov/ Identifier: NCT02275637.
2020
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3366041
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 24
  • ???jsp.display-item.citation.isi??? 30
social impact