Among the implantable cardioverter defibrillator recipients, there is still a subgroup of patients in whom the defibrillation threshold is too high and the maximal shock output of the implantable cardioverter defibrillator can fail to terminate a ventricular arrhythmia. We report a new thoracoscopic minimally invasive approach to place a standard array electrode in the transverse pericardial sinus of a patient implanted with a cardiac resynchronization and defibrillation therapy device with persistent high defibrillation threshold. This approach was developed to achieve very low shock impedance with a consequent increase in the current flow and reduction of defibrillation threshold.
Thoracoscopic Implantation of an Array Electrode in the Pericardium Transverse Sinus to Reduce Defibrillation Threshold
Giacopelli D.;
2017
Abstract
Among the implantable cardioverter defibrillator recipients, there is still a subgroup of patients in whom the defibrillation threshold is too high and the maximal shock output of the implantable cardioverter defibrillator can fail to terminate a ventricular arrhythmia. We report a new thoracoscopic minimally invasive approach to place a standard array electrode in the transverse pericardial sinus of a patient implanted with a cardiac resynchronization and defibrillation therapy device with persistent high defibrillation threshold. This approach was developed to achieve very low shock impedance with a consequent increase in the current flow and reduction of defibrillation threshold.Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.