Objective: To evaluate the oxygen reserve index (ORI) as a noninvasive estimate of the PaO2 during moderate hyperoxaemia [100–200 mmHg (13.3–26.6 kPa)], and to determine ORI values identifying PaO2 > 100, > 150 (20.0 kPa) and > 200 mmHg in anaesthetized donkeys with an inspired fraction of oxygen (FIO2) > 0.95. Study design: Prospective observational study. Animals: A group of 28 adult standard donkeys aged (mean ± standard deviation) 4 ± 2 years and weighing 135 ± 15 kg. Methods: Donkeys were sedated intramuscularly with xylazine and butorphanol; anaesthesia was induced with ketamine and diazepam and maintained with isoflurane in oxygen. An adhesive sensor probe was applied to the donkey's tongue and connected to a Masimo pulse co-oximeter to determine ORI values. An arterial catheter was inserted into an auricular artery. After ORI signal stabilization, the value was noted and PaO2 determined by blood gas analysis. The Pearson correlation coefficient was used to assess the relationship between ORI and PaO2 for oxygen tension < 200 mmHg (< 26.6 kPa). The Youden index was used to identify the value of ORI that detected PaO2 > 150 and 200 mmHg (20.0 and 26.6 kPa) with the highest sensitivity and specificity. Results: A total of 106 paired measurements were collected. A mild positive correlation was observed between ORI and PaO2 for values < 200 mmHg (26.6 kPa; r = 0.52). An ORI > 0.0, > 0.1 and > 0.3 indicated a PaO2 > 100, > 150 and > 200 mmHg (13.3, 20.0 and 26.6 kPa) with negative predictive values > 94%. Conclusions and clinical relevance: ORI may provide a noninvasive indication of PaO2 > 100, > 150 and > 200 mmHg (13.3, 20.0 and 26.6 kPa) in anaesthetized donkeys with an FIO2 > 0.95, although it does not replace blood gas analysis for assessment of oxygenation.

Oxygen reserve index as a noninvasive indicator of arterial partial pressure of oxygen in anaesthetized donkeys: a preliminary study

Bellini L.
;
De Benedictis G. M.;
2021

Abstract

Objective: To evaluate the oxygen reserve index (ORI) as a noninvasive estimate of the PaO2 during moderate hyperoxaemia [100–200 mmHg (13.3–26.6 kPa)], and to determine ORI values identifying PaO2 > 100, > 150 (20.0 kPa) and > 200 mmHg in anaesthetized donkeys with an inspired fraction of oxygen (FIO2) > 0.95. Study design: Prospective observational study. Animals: A group of 28 adult standard donkeys aged (mean ± standard deviation) 4 ± 2 years and weighing 135 ± 15 kg. Methods: Donkeys were sedated intramuscularly with xylazine and butorphanol; anaesthesia was induced with ketamine and diazepam and maintained with isoflurane in oxygen. An adhesive sensor probe was applied to the donkey's tongue and connected to a Masimo pulse co-oximeter to determine ORI values. An arterial catheter was inserted into an auricular artery. After ORI signal stabilization, the value was noted and PaO2 determined by blood gas analysis. The Pearson correlation coefficient was used to assess the relationship between ORI and PaO2 for oxygen tension < 200 mmHg (< 26.6 kPa). The Youden index was used to identify the value of ORI that detected PaO2 > 150 and 200 mmHg (20.0 and 26.6 kPa) with the highest sensitivity and specificity. Results: A total of 106 paired measurements were collected. A mild positive correlation was observed between ORI and PaO2 for values < 200 mmHg (26.6 kPa; r = 0.52). An ORI > 0.0, > 0.1 and > 0.3 indicated a PaO2 > 100, > 150 and > 200 mmHg (13.3, 20.0 and 26.6 kPa) with negative predictive values > 94%. Conclusions and clinical relevance: ORI may provide a noninvasive indication of PaO2 > 100, > 150 and > 200 mmHg (13.3, 20.0 and 26.6 kPa) in anaesthetized donkeys with an FIO2 > 0.95, although it does not replace blood gas analysis for assessment of oxygenation.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3400990
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact