Background: Reliable post-approval surveillance of implantable cardioverter-defibrillator (ICD) lead performance remains a challenge. In the past, two ICD leads were recalled due to a high frequency of failures. In this meta-analysis, we sought to provide a combined estimate of failure-free rate for ICD leads by reconstructing individual patient data from published Kaplan-Meier (KM) curves and to investigate whether estimates could be influenced by the characteristics of the study. Methods: Observational studies assessing failure-free estimates of transvenous ICD leads with KM method, were identified through a systematic search up to November 2021. Results: Forty-four studies were eligible that included 41,870 (63.1%) non-recalled leads and 24,493 (36.9%) recalled leads. The 8-year cumulative failure-free rate was 94.1% (CI, 93.6% - 94.6%) for contemporary non-recalled leads and 81.2% (80.3% - 82.0%) for recalled leads (hazard ratio [HR], 3.15 [2.85-3.47], p < 0.001). Failure-free rate was lower in single-center studies in both the non-recalled (HR, 0.28 [0.15-0.51], p < 0.001) and recalled (HR, 0.54 [0.33-0.88], p = 0.014) group compared with multicenter studies. Similarly, estimates were significantly lower in small (i.e. extracted KM curve with <312 leads) versus large studies (HR non-recalled group, 0.54 [CI, 0.33-0.89], p = 0.015; HR recalled group, 0.62 [CI, 0.43-0.89], p = 0.009). Conclusions: In this meta-analysis including >66,000 leads, we provide pooled survival curves that may play a role in generating evidence-based standards for assessing clinically acceptable failure rates for ICD leads. Lead performance was underestimated with single-center and small-sized studies; multicenter studies remain the main tool to reliably conduct post-market surveillance of ICD leads.

Implantable cardioverter defibrillator lead performance: A systematic review and individual patient data Meta-analysis

Giacopelli D.;Azzolina D.;Gregori D.
2023

Abstract

Background: Reliable post-approval surveillance of implantable cardioverter-defibrillator (ICD) lead performance remains a challenge. In the past, two ICD leads were recalled due to a high frequency of failures. In this meta-analysis, we sought to provide a combined estimate of failure-free rate for ICD leads by reconstructing individual patient data from published Kaplan-Meier (KM) curves and to investigate whether estimates could be influenced by the characteristics of the study. Methods: Observational studies assessing failure-free estimates of transvenous ICD leads with KM method, were identified through a systematic search up to November 2021. Results: Forty-four studies were eligible that included 41,870 (63.1%) non-recalled leads and 24,493 (36.9%) recalled leads. The 8-year cumulative failure-free rate was 94.1% (CI, 93.6% - 94.6%) for contemporary non-recalled leads and 81.2% (80.3% - 82.0%) for recalled leads (hazard ratio [HR], 3.15 [2.85-3.47], p < 0.001). Failure-free rate was lower in single-center studies in both the non-recalled (HR, 0.28 [0.15-0.51], p < 0.001) and recalled (HR, 0.54 [0.33-0.88], p = 0.014) group compared with multicenter studies. Similarly, estimates were significantly lower in small (i.e. extracted KM curve with <312 leads) versus large studies (HR non-recalled group, 0.54 [CI, 0.33-0.89], p = 0.015; HR recalled group, 0.62 [CI, 0.43-0.89], p = 0.009). Conclusions: In this meta-analysis including >66,000 leads, we provide pooled survival curves that may play a role in generating evidence-based standards for assessing clinically acceptable failure rates for ICD leads. Lead performance was underestimated with single-center and small-sized studies; multicenter studies remain the main tool to reliably conduct post-market surveillance of ICD leads.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3474574
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
social impact