The validation process is essential in accredited laboratory medicine, but is rarely regarded as an issue in the preanalytical management. The aim of this study was to validate five kinds of lithium heparin vacuum tubes for routine clinical chemistry laboratory testing. Blood specimens from 100 volunteers in five different plasma vacuum tubes (Tube I: VACUETTE®, Tube II: LABOR IMPORT®, Tube III: S-Monovette®, Tube IV: PST® and Tube V: PST II®) were collected by a single expert phlebotomist. The routine clinical chemistry tests were performed on a Cobas® 6000 <c501> module. The significance of the differences between samples was statistically assessed at p < 0.005. The biases from the different tubes were compared with the current desirable quality specifications. Basically, significant differences could be confirmed by RM ANOVA for the results of the clinical chemistry tests on the following components: glucose, urea, creatinine, alkaline phosphatase, amylase, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, total bilirubin, phosphate, Ca, Mg, Fe and K. Clinically significant variations as compared with the current desirable quality specifications were found for glucose, creatinine, amylase, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, Ca, Mg and K. In conclusion, our results do not support arbitrary interchange among brands of plasma vacuum tubes. Future investigations are needed to understand the reasons of these observations; in the meantime, we suggest that laboratory managers standardize the procedures and frequently evaluate the quality of in vitro diagnostic devices.
Quality management of preanalytical phase: Impact of lithium heparin vacuum tubes changes on clinical chemistry tests
The validation process is essential in accredited laboratory medicine, but is rarely regarded as an issue in the preanalytical management. The aim of this study was to validate five kinds of lithium heparin vacuum tubes for routine clinical chemistry laboratory testing. Blood specimens from 100 volunteers in five different plasma vacuum tubes (Tube I: VACUETTE®, Tube II: LABOR IMPORT®, Tube III: S-Monovette®, Tube IV: PST® and Tube V: PST II®) were collected by a single expert phlebotomist. The routine clinical chemistry tests were performed on a Cobas® 6000 module. The significance of the differences between samples was statistically assessed at p < 0.005. The biases from the different tubes were compared with the current desirable quality specifications. Basically, significant differences could be confirmed by RM ANOVA for the results of the clinical chemistry tests on the following components: glucose, urea, creatinine, alkaline phosphatase, amylase, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, total bilirubin, phosphate, Ca, Mg, Fe and K. Clinically significant variations as compared with the current desirable quality specifications were found for glucose, creatinine, amylase, aspartate aminotransferase, alanine aminotransferase, lactate dehydrogenase, Ca, Mg and K. In conclusion, our results do not support arbitrary interchange among brands of plasma vacuum tubes. Future investigations are needed to understand the reasons of these observations; in the meantime, we suggest that laboratory managers standardize the procedures and frequently evaluate the quality of in vitro diagnostic devices.
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.
Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3538655
Citazioni
ND
10
8
ND
social impact
Conferma cancellazione
Sei sicuro che questo prodotto debba essere cancellato?
simulazione ASN
Il report seguente simula gli indicatori relativi alla propria produzione scientifica in relazione alle soglie ASN 2023-2025 del proprio SC/SSD. Si ricorda che il superamento dei valori soglia (almeno 2 su 3) è requisito necessario ma non sufficiente al conseguimento dell'abilitazione. La simulazione si basa sui dati IRIS e sugli indicatori bibliometrici alla data indicata e non tiene conto di eventuali periodi di congedo obbligatorio, che in sede di domanda ASN danno diritto a incrementi percentuali dei valori. La simulazione può differire dall'esito di un’eventuale domanda ASN sia per errori di catalogazione e/o dati mancanti in IRIS, sia per la variabilità dei dati bibliometrici nel tempo. Si consideri che Anvur calcola i valori degli indicatori all'ultima data utile per la presentazione delle domande.
La presente simulazione è stata realizzata sulla base delle specifiche raccolte sul tavolo ER del Focus Group IRIS coordinato dall’Università di Modena e Reggio Emilia e delle regole riportate nel DM 589/2018 e allegata Tabella A. Cineca, l’Università di Modena e Reggio Emilia e il Focus Group IRIS non si assumono alcuna responsabilità in merito all’uso che il diretto interessato o terzi faranno della simulazione. Si specifica inoltre che la simulazione contiene calcoli effettuati con dati e algoritmi di pubblico dominio e deve quindi essere considerata come un mero ausilio al calcolo svolgibile manualmente o con strumenti equivalenti.