A comparative study of the neutron-γ discrimination performance of a liquid scintillator detector BC501A coupled to four different 5 in. photomultiplier tubes (ET9390kb, R11833-100, XP4512 and R4144) was carried out. Both the Charge Comparison method and the Integrated Rise-Time method were implemented digitally to discriminate between neutrons and γ rays emitted by a 252Cf source. In both methods, the neutron-γ discrimination capabilities of the four photomultiplier tubes were quantitatively compared by evaluating their figure-of-merit values at different energy regions between 50 keVee and 1000 keVee. Additionally, the results were further verified qualitatively using time-of-flight to distinguish γ rays and neutrons. The results consistently show that photomultiplier tubes R11833-100 and ET9390kb generally perform best regarding neutron-γ discrimination with only slight differences in figure-of-merit values. This superiority can be explained by their relatively higher photoelectron yield, which indicates that a scintillator detector coupled to a photomultiplier tube with higher photoelectron yield tends to result in better neutron-γ discrimination performance. The results of this work will provide reference for the choice of photomultiplier tubes for future neutron detector arrays like NEDA.
Test of digital neutron-gamma discrimination with four different photomultiplier tubes for the NEutron Detector Array (NEDA)
Triossi A.;
2014
Abstract
A comparative study of the neutron-γ discrimination performance of a liquid scintillator detector BC501A coupled to four different 5 in. photomultiplier tubes (ET9390kb, R11833-100, XP4512 and R4144) was carried out. Both the Charge Comparison method and the Integrated Rise-Time method were implemented digitally to discriminate between neutrons and γ rays emitted by a 252Cf source. In both methods, the neutron-γ discrimination capabilities of the four photomultiplier tubes were quantitatively compared by evaluating their figure-of-merit values at different energy regions between 50 keVee and 1000 keVee. Additionally, the results were further verified qualitatively using time-of-flight to distinguish γ rays and neutrons. The results consistently show that photomultiplier tubes R11833-100 and ET9390kb generally perform best regarding neutron-γ discrimination with only slight differences in figure-of-merit values. This superiority can be explained by their relatively higher photoelectron yield, which indicates that a scintillator detector coupled to a photomultiplier tube with higher photoelectron yield tends to result in better neutron-γ discrimination performance. The results of this work will provide reference for the choice of photomultiplier tubes for future neutron detector arrays like NEDA.Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.




