During percutaneous microcompression of the gasserian ganglion for the relief of trigeminal neuralgia, a computerized technique for monitoring the pressure inside Meckel's cave was employed in 22 patients. A dedicated transducer connected to a computer records the balloon inflation pressure. Its variations are discernible within tenths of a bar and are plotted in relation to time. The intraoperative pressure inside Meckel's cave is from 0.9 to 2.4 bars. When pressure was low, there was recurrence of pain. The highest values of pressure (1.9-2.4 bars) were observed in most of the patients suffering from untoward side effects. The clinical results seem to be influenced by the level of the intraoperative intracavitary pressure.

Pressure monitoring inside Meckel's cave during percutaneous microcompression of gasserian ganglion

Landi, A;
1991

Abstract

During percutaneous microcompression of the gasserian ganglion for the relief of trigeminal neuralgia, a computerized technique for monitoring the pressure inside Meckel's cave was employed in 22 patients. A dedicated transducer connected to a computer records the balloon inflation pressure. Its variations are discernible within tenths of a bar and are plotted in relation to time. The intraoperative pressure inside Meckel's cave is from 0.9 to 2.4 bars. When pressure was low, there was recurrence of pain. The highest values of pressure (1.9-2.4 bars) were observed in most of the patients suffering from untoward side effects. The clinical results seem to be influenced by the level of the intraoperative intracavitary pressure.
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3301805
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 17
  • ???jsp.display-item.citation.isi??? 12
social impact