Aim: Whether glucose sensor alarms improve metabolic control and are accepted by individuals with diabetes is unclear. Here, we investigated whether switching from a standard flash glucose monitoring system (FGM1) to a system equipped with hypo- and hyperglycemia alarms (FGM2) improves glycemic control and psychological outcomes in adults with type 1 diabetes (T1D). Methods: Subjects with T1D and > 4% of time in hypoglycemia or > 40% of time in hyperglycemia were studied while wearing FGM1 (4 weeks) and after switching to FGM2 for 8 weeks. The primary endpoint was the change in time in range (TIR 70–180 mg/dl [3.9–10.0 mmol/L]) after 4 weeks of FGM2 use. Time below range (TBR), time above range (TAR), mean glucose, coefficient of variation (CV), sensor scans, treatment satisfaction, and hypoglycemia fear were secondary outcomes. Results: We included 38 subjects aged 33.7 ± 12.6 year. During 4 weeks of FGM2 use, TIR increased from 52.8 to 57.0% (p = 0.001), TBR decreased from 6.2 to 3.4% (p < 0.0001) as did time < 54 mg/dl (from 1.4 to 0.3%, p < 0.0001) and CV (from 39.6% to 36.1%, p < 0.0001). These changes were confirmed after 8 weeks of FGM2 use. Treatment satisfaction improved and fear of hypoglycemia decreased. Subjects who had > 4% of time in hypoglycemia at baseline showed the greatest improvements in glucose control and treatment satisfaction. Conclusion: Switching from FGM1 to FGM2 improved TIR and treatment satisfaction and reduced fear of hypoglycemia. Participants who benefited most from switching from FGM1 to FGM2 were those prone to hypoglycemia.

Effectiveness of adding alarms to flash glucose monitoring in adults with type 1 diabetes under routine care

Boscari F.;Ferretto S.;Fadini G. P.;Avogaro A.;Bruttomesso D.
2022

Abstract

Aim: Whether glucose sensor alarms improve metabolic control and are accepted by individuals with diabetes is unclear. Here, we investigated whether switching from a standard flash glucose monitoring system (FGM1) to a system equipped with hypo- and hyperglycemia alarms (FGM2) improves glycemic control and psychological outcomes in adults with type 1 diabetes (T1D). Methods: Subjects with T1D and > 4% of time in hypoglycemia or > 40% of time in hyperglycemia were studied while wearing FGM1 (4 weeks) and after switching to FGM2 for 8 weeks. The primary endpoint was the change in time in range (TIR 70–180 mg/dl [3.9–10.0 mmol/L]) after 4 weeks of FGM2 use. Time below range (TBR), time above range (TAR), mean glucose, coefficient of variation (CV), sensor scans, treatment satisfaction, and hypoglycemia fear were secondary outcomes. Results: We included 38 subjects aged 33.7 ± 12.6 year. During 4 weeks of FGM2 use, TIR increased from 52.8 to 57.0% (p = 0.001), TBR decreased from 6.2 to 3.4% (p < 0.0001) as did time < 54 mg/dl (from 1.4 to 0.3%, p < 0.0001) and CV (from 39.6% to 36.1%, p < 0.0001). These changes were confirmed after 8 weeks of FGM2 use. Treatment satisfaction improved and fear of hypoglycemia decreased. Subjects who had > 4% of time in hypoglycemia at baseline showed the greatest improvements in glucose control and treatment satisfaction. Conclusion: Switching from FGM1 to FGM2 improved TIR and treatment satisfaction and reduced fear of hypoglycemia. Participants who benefited most from switching from FGM1 to FGM2 were those prone to hypoglycemia.
2022
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11577/3451486
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 8
  • OpenAlex ND